Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
TECHNICAL NOTE
Comparison of Brain Volume Measurements Made with 0.3- and 3-T MR Imaging
Syo Murata, Akifumi Hagiwara, Hideyoshi Kaga, Yuki Someya, Kiyotaka Nemoto, Masami Goto, Koji Kamagata, Ryusuke Irie, Masaaki Hori, Christina Andica, Akihiko Wada, Kanako Kunishima Kumamaru, Keigo Shimoji, Yujiro Otsuka, Haruyoshi Hoshito, Yoshifumi Tamura, Ryuzo Kawamori, Hirotaka Watada, Shigeki Aoki
Author information
JOURNAL OPEN ACCESS
Supplementary material

2022 Volume 21 Issue 3 Pages 517-524

Details
Abstract

The volumes of intracranial tissues of 40 healthy volunteers acquired from 0.3- and 3-T scanners were compared using intraclass correlation coefficients, correlation analyses, and Bland-Altman analyses. We found high intraclass correlation coefficients, high Pearson’s correlation coefficients, and low percentage biases in all tissues and most of the brain regions, although small differences were observed in some areas. These findings may support the validity of brain volumetry with low-field magnetic resonance imaging.