Magnetic Resonance in Medical Sciences
Online ISSN : 1880-2206
Print ISSN : 1347-3182
ISSN-L : 1347-3182
Technical Note
Optimized System Design and Construction of a Compact Whole-hand Scanner for Diagnosis of Rheumatoid Arthritis
Shinya HANDAHiroshi YOSHIOKASadanori TOMIHATomoyuki HAISHIKatsumi KOSE
Author information
JOURNAL OPEN ACCESS

2007 Volume 6 Issue 2 Pages 113-120

Details
Abstract

We have developed a compact magnetic resonance (MR) imaging scanner with permanent magnet, gradient coil set, and radiofrequency (RF) coils optimized for whole-hand examination for the diagnosis of rheumatoid arthritis (RA). The system weighs about 600 kg, and installation space is 2 m2, excluding the shield room. Hand examinations of normal volunteers and patients with RA were performed using a 3D T1-weighted gradient-echo (GRE) sequence and short TI inversion recovery 3D fast spin-echo (STIR-3DFSE) sequence, and anatomical structures and various lesions of the hand caused by RA were clearly visualized in a 16-min examination. It was concluded that the system could be used for diagnosis of RA in even a small clinic.

Content from these authors
© 2007 by Japanese Society for Magnetic Resonance in Medicine
Previous article Next article
feedback