Three Dimensional Ultra-short Echo Time MRI Can Depict Cholesterol Components of Gallstones Bright
Mamoru TakahashiYasuo TakeharaKenji FujisakiTomoyuki OkuakiYukiko FukumaNorihiro TooyamaKatsutoshi IchijoTomoyasu AmanoSatoshi GoshimaShinji Naganawa
Author information
JOURNAL OPEN ACCESS Advance online publication

Article ID: mp.2020-0009

Details
Abstract

Purpose: Non-calcified cholesterol stones that are small in size are hard to be depicted on CT or magnetic resonance cholangiopancreatography. This institutional review board (IRB)-approved retrospective in vitro study aims to characterize contrast behaviors of 3 main components of the gallstones, i.e., cholesterol component (CC), bilirubin calcium component (BC) and CaCO3 (CO) on 3D radial scan with ultrashort TE (UTE) MRI, and to test the capability of depicting CC of gallstones as bright signals as compared to background saline.

Methods: Fourteen representative gallstones from 14 patients, including 15 CC, 6 BC and 4 CO, were enrolled. The gallstones underwent MRI including fat-saturated T1-weighted image (fs-T1WI) and UTE MRI with dual echoes. The contrast-to-noise ratio (CNR) and the chemical analysis for the 25 portions of the stones were compared.

Results: BC was bright on fs-T1WI, which did not change dramatically on UTE MRI and the signal did not remain on UTE subtraction image between dual echoes. Whereas the CC was negative or faintly positive signal on fs-T1WI, bright signal on UTE MRI and the contrast remained even higher on the UTE subtraction, which reflected their short T2 values. Median CNRs and standard errors of the segments on each imaging were as follows: on fs-T1WI, −10.2 ± 4.2 for CC, 149.7 ± 27.6 for BC and 37.9 ± 14.3 for CO; on UTE MRI first echo, 16.7 ± 3.3 for CC, 74.9 ± 21.3 for BC and 17.7 ± 8.4 for CO; on UTE subtraction image, 30.2 ±2.0 for CC, −11.2 ± 5.4 for BC and 17.8 ± 10.7 for CO. Linear correlations between CNRs and cholesterol concentrations were observed on fs-T1WI with r = −0.885, (P < 0.0001), UTE MRI first echo r = −0.524 (P = 0.0072) and UTE subtraction with r = 0.598 (P = 0.0016).

Conclusion: UTE MRI and UTE subtraction can depict CC bright.