Neurosonology
Print ISSN : 0917-074X
ISSN-L : 0917-074X
Development and Clinical Usefulness of an Ultrasonic Aspirator for a Stereotactic Neuro-endoscopic Surgical System
Kunihiko EBINAShigeharu SUZUKI
Author information
JOURNAL FREE ACCESS

1996 Volume 9 Issue 3 Pages 85-90

Details
Abstract

We have developed an originally designed ultrasonic surgical aspirator unit for a newly developed stereotactic neuro-endoscopic surgical system. It has a maximum power output of 35 W, maximum amplitude of 90 μm, a resonance frequency of 23 kHz, capability for fine adjustment using 12 levels, and oscillating probes prepared in three sizes, 2.0×1.2×310 mm (external diameter × internal diameter × effective length), 2.0×1.3×310 mm, and 1.8×1.1×295 mm.
This unit fragments and aspirates hard blood clots, viscous fluid and tumors, thus facilitating efficient CT-guided stereotactic neuro-endoscopic brain surgery. We have applied this unit to 40 cases of intracerebral hematoma, 26 cases of brain tumor, and 5 other cases. This unit appears to be less invasive and safer, as it can be applied to removal of not only intracerebral hematomas but also deep-seated brain tumors, and also neuronal transplantation.

Content from these authors
© The Japan Academy of Neurosonology
Previous article Next article
feedback
Top