NIPPON KAGAKU KAISHI
Online ISSN : 2185-0925
Print ISSN : 0369-4577
The Simplified Determination of Magnesium in Water by Calcium Carbonate Coprecipitation-Flame Atomic Absorption Spectrometry
Futoshi SUGIMOTOTeizo ISONOKazuhiko KOMURASAKI
Author information
JOURNAL FREE ACCESS

2002 Volume 2002 Issue 2 Pages 175-181

Details
Abstract
The calcium carbonate coprecipitation method was studied for simplified determination of magnesium ions in water by flame atomic absorption spectrometry (F-AAS).
A 15 mL of 8 gCa L−1 calcium chloride and 15 mL of 20 w/v% potassium carbonate as a coprecipitant were added to sample water containing magnesium ions, followed by stirring for 30 min at room temperature. The resulting magnesium ions were quantitatively coprecipitated with calcium carbonate. The precipitate was collected onto a membrane filter (0.45 μm) and then dissolved with dipping for 15 min in 10 mL of (1 + 11)HCl. The solution was made up to 100 mL with distilled water to prepare the sample solution for F-AAS.
The detection limit for magnesium was 0.9 μg L−1(3σ), and the lower limit of determination was 3.2 μg L−1(10σ) in case of 200 mL of sample water.
This calcium carbonate coprecipitation method has no significant interferences in the analysis of water samples of 50 to 400 mL, and could be used as a preconcentration method for simplified determination by F-AAS of magnesium ions in water.
Content from these authors

This article cannot obtain the latest cited-by information.

© 2002 The Chemical Society of Japan
Previous article Next article
feedback
Top