1981 Volume 1981 Issue 4 Pages 616-618
Use of silver salt was examined to enhance reliability of the flow injection analysis (FIA) for the chemical oxygen demand with potassium permanganate (CODFIA⋅Mn). To compare with CODm based on a Japan Industrial Standard, pertinent operating conditions of the FIA were investigated for waste water samples by using D-glucose or lactose as a standard. It has turned out that the most reliable results could be obtained by adding O.05% of silver nitrate into the flow of acidic permanganate solution as a catalyst. Under the proposed conditions, chloride up to 1000 mg⋅l-1 did not interfere with the determination. The newly proposed FIA method could be successfully applied to industrial waste water samples.