NIPPON KAGAKU KAISHI
Online ISSN : 2185-0925
Print ISSN : 0369-4577
Use of Column-Switching Technique in Photometric Ion Chromatography
Ikuyo YOSHIDAKazuichi HAYAKAWAMotoichi MIYAZAKI
Author information
JOURNAL FREE ACCESS

1986 Volume 1986 Issue 7 Pages 1046-1049

Details
Abstract

The effectiveness of column-switching technique in photometric ion chromatography was investigated by use of two columns, into which resins with different specific ion-exchange capacity were packed. A system is consisted of two pumps, a sample injector, two columns, a switching valve, two UV detectors and a dual-pen recorder. As a result, it was found that the column-switching technique possessed the following advantages: (1)rapid and simultaneous determination pf, yarious ions with noticeably different capacity ratios; (2) excellent separation of the peak which can not be resolved by use of a single column; (3) elimination of system peak.

Content from these authors

This article cannot obtain the latest cited-by information.

© The Chemical Society of Japan
Previous article
feedback
Top