Characteristics of Sodium Ion-Selective Electrodes Based on Neutral Carriers and Their Application to Analysis of Physiological Fluids
Yasuhisa SHIBATATakuya MARUIZUMIHiroyuki MIYAGI
Author information
JOURNAL FREE ACCESS

1992 Volume 1992 Issue 9 Pages 961-967

Details
Abstract

Characteristics of sodium ion-selective electrodes (Na-ISEs) with polymeric membranes incorporating cyclic and acyclic compounds as ionophores were studied and their applicability to analysis of physiological fluids was examined. In this study, bis[(12-crown-4)methyl]methyl(dodecyl)malonate [bis(12-crown-4)] and polyether-amide com pounds (ETH 157, ETH 2120) were used as ionophores and adipic esters [bis(2-ethylhexyl) adipate, bis(1-butylpentyl)adipate] were used as plasticizers. These Na-ISEs showed Nernstian slope of 54-57 mV/decade and response time of <10 s. Among three ionophores examined, bis(12-crown-4) was the most selective to Na + in the presence of drugs [1, 1-dimethy1-5-methoxy-3-(di-2-thienylmethylene)piperidinium bromide monohydrate and [2-(α, α-dicyclopentyl acetoxy)ethyl]triethylammonium bromide] and lipophilic quaternary ammonium salts such as a cationic surfactant (trimethyloctadecylammonium salt) as well as cations. The cause of interference to three Na-ISEs by the above two drugs was studied. Then, the migration of interfering substances into the sensing membrane was suggested by the measurement of potential response characteristics. In the analysis of control serum samples, good reproducibility (C. V. =0.3-0.4%) was obtained by all Na-ISEs. But accuracy in the analysis of patient serum samples was varied with the type of ionophore of Na-ISE. The deviation of error (Syx) was as follows,
ETH 157>ETH 2120>bis(12-crown-4)
Therefore, Na-ISE based on bis(12-cro wn-4) was applied to the analysis of serum, plasma and urine samples. Good correlation was obtained between the results of this Na-ISE and flame photometry. From the point of stability of Na-ISE and accuracy in the analysis of physiological fluids, it was suggested that the Na-ISE based on bis(12-crown-4) as an ionophore would be useful in the field of clinical analysis.

Content from these authors

This article cannot obtain the latest cited-by information.

© The Chemical Society of Japan
Previous article Next article
feedback
Top