PDA Journal of GMP and Validation in Japan
Online ISSN : 1881-1728
Print ISSN : 1344-4891
ISSN-L : 1344-4891
Topics Review
Utilization of the Steam BIER (Biological Indicator Evaluator Resistometer) Vessel on Sterility Assurance for Pharmaceuticals—Application to Sterilization Process Validation for Parenteral Drugs
Takuji KAMOGAWA
Author information
JOURNAL FREE ACCESS

2002 Volume 4 Issue 2 Pages 99-106

Details
Abstract

  In the development or validation of sterilization process for terminally sterilized parenteral drugs, biological indicators (BIs) are used to evaluate the effectiveness of sterilization. Accurately determining the heat resistance of BIs (D value) in parenrteral solution is extremely important because the D value varies due to variations in the chemical composition of parenteral solutions. The steam BIER vessel can rigidly control the process variables such as exposure temperature and time and provide reproducible physical test conditions. It is possible to heat the BI suspension in capillary tube or small container with short time for come-up and cool-down by using the steam BIER vessel. Because of this advantage, the steam BIER vessel is useful for determination of the D value in parenteral solution. The steam BIER vessel can solve the problem concerning accuracy and procedures of determining the D value in parenteral solution and contribute to perform the high level sterilization process validation for parenteral drugs.

Content from these authors
© 2002 Parenteral Drug Association Japan Chapter
Previous article Next article
feedback
Top