Journal of The Japanese Society for Quality Control
Online ISSN : 2432-1044
Print ISSN : 0386-8230
Quality Report
Optimization of the Calibration Period in the Case that“JIS Z 9090” is not Applicable-In the Case of Pin Gauges-
Ichiro SASAKIWenjing GAOXifeng WANG
Author information
JOURNAL RESTRICTED ACCESS

2016 Volume 46 Issue 4 Pages 421-426

Details
Abstract
 “JIS Z 9090 :Measurement-General rules for calibration sistem―Appendix1 : How to determine the calibration period and the correction limit”is a standard to optimize the calibration period of measurement systems.It’s based on the method of Feedback Control in On-line Quality Engineering. But pingauges, etc. are not applicable to this standard. Therefore we examined another methods applicable in that case.After consideration,our conclusion is that the method of Diagnosis and Adjustment is appropriate. And we took the concept calculating the costs by quantifying the risk under a variety of circumstances of calibration.As a result,the calibration period of pingauges in a china plant could be extended to three years from one year.At the same time,the calibration cost could be reduced to one third.
Content from these authors
© 2016 The Japanese Society for Quality Control