2018 Volume 7 Issue 1 Pages 53-57
For the accurate calibration of CR-39 based SSNTDs, a protocol shall be given which contains parameters of the calibration chamber as well as the calibration and reference method. The specification of the reference method is an important task. During this work various gamma spectrometric and scintillation techniques capable for thoron measurement were tested in the interest of selecting a reference method for the validation of thoron calibration. Comparison measurements were performed in the calibration chamber with the suitable techniques and the available active devices. Based on the results, scintillation counting according to two counts method has been proposed as reference method for thoron measurements.