Radiation Safety Management
Online ISSN : 1884-9520
Print ISSN : 1347-1511
ISSN-L : 1347-1511
Papers
Development of Faced-Type Double NaI(Tl) Scintillation Detector System for Measurement of Radioactive-Cesium Concentration
Yuki YANONaoki ZUSHIFuminobu SATOYushi KATOToshiyuki IIDA
Author information
JOURNAL FREE ACCESS

2015 Volume 14 Issue 1 Pages 1-8

Details
Abstract
  A low-cost and high-performance gamma-ray spectrometer has been successfully constructed by using 3″φ × 3″ NaI(Tl) scintillation probes, standard lead bricks and others for the measurement of radioactive-cesium concentration of various surrounding materials. A faced double detector system with signal mixing has been employed for the improvement of total-absorption detection efficiency for radioactive-cesium. Uniquely-shaped containers covering the double detectors have been made of acrylic pipes and plates. The total-absorption detection efficiency for radioactive-cesium in the homemade container was experimentally determined by using a standard voluminous source made of soil in Fukushima. The detection efficiency determined from the experiments almost agreed results from Monte Carlo calculations with the PHITS code. It was confirmed from radioactivecesium measurement on fly ash that the subtraction of energy spectral data of background radiations was useful for low-level radioactive-cesium measurement.
Content from these authors
© 2015 Japanese Society of Radiation Safety Managenent
Next article
feedback
Top