Radiation Safety Management
Online ISSN : 1884-9520
Print ISSN : 1347-1511
ISSN-L : 1347-1511
High Sensitive123I Thyroid Uptake Measurement Method Using an Imaging Plate
Shigeki ItoTakuya SazeKunihide Nishizawa
Author information
JOURNAL FREE ACCESS

2004 Volume 3 Issue 1 Pages 11-19

Details
Abstract

A high sensitive 123I thyroid uptake measurement method was devised by using an imaging plate and a bio imaginganalyzer system (IP system).The IP system gives an image reflecting thyroid outline.The IP system can detectthe thyroid activity of 600 Bq corresponding to 5% uptake 24 hr after administration of 42 kBq within an uncertaintyof 5%.The thyroid uptake study using IP system can be conducted by administrating the 1.1% activity of 3.7 MBqrequired for the NaI system in accordance with the guideline recommended by The Society of Nuclear Medicine.Thethyroid-absorbed dose for a normal adult administered 42 kBq was 0.13 mGy, that is comparable to the equivalentdose of the thyroid of 0.1 mGy due to the chest X-ray examination at the guidance level of 0.4 mGy recommended by IAEA.The PSL value-activity conversion coefficient was 6.0×10-3±0.15×10-3 PSL.sec-1.Bq-1.

Content from these authors
© Japanese Society of Radiation Safety Management
Previous article
feedback
Top