SEIBUTSU BUTSURI KAGAKU
Online ISSN : 1349-9785
Print ISSN : 0031-9082
ISSN-L : 0031-9082
Workshop
Clinical laboratory evaluation of SELECA-VSP for serum protein fractionation by use of automated electrophoresis system (AES320)
Hiroko WatanabeAya SugiuraMasato Maekawa
Author information
JOURNAL FREE ACCESS

2007 Volume 51 Issue 2 Pages 125-128

Details
Abstract

Serum protein fractionation has been performed by electrophoresis on cellulose acetate membrane in clinical laboratory and has been applied to diagnostic procedure for pathological states. In Japan SEPARAX-SP (Fuji film Co.) has been used as the supporting media of cellulose acetate membrane for about twenty years and will finish its production soon. A new cellulose acetate membrane, SELECA-VSP (Toyo Roshi Kaisha Ltd.) has been developed instead of SEPARAX-SP. The SELECA-VSP has almost similar features and function. In this study we evaluated quality and utility of SELECA-VSP by automated electrophoresis system (Olympus AES320). Protein fractionation using SELECA-VSP was not worse than that of SEPARAX-SP. Separation sensitivity was higher in SELECA-VSP. However, irregularly smeared bands were sometimes observed adjacent to monoclonal protein bands. Areas of each fraction are highly correlated in between SEPARAX-SP and SELECA-VSP. The fraction area of α1-globulin was significantly different in SEPARAX-SP and SELECA-VSP, and about 0.01 g / dl higher in SELECA-VSP. After the presence of the difference is recognized, automated classification system for pathological states by protein fractionation could be utilized for routine laboratory testing.
In conclusion, the new cellulose acetate membrane, SELECA-VSP will replace SEPARAX-SP as electrophoretic supporting media for serum protein fractionation.

Content from these authors
© 2007 by Japanese Electrophoresis Society
Previous article Next article
feedback
Top