Host: The Society of Instrument and Control Engineers
Co-host: IEEE/Industiral Electronic Society, IEEE/Robotics and Automation Society, IEEE/Control System Society
Pages 227
NiC eutectic point (approximately 1329°C) is studied as a fixed point for the calibration of Pt/Pd thermocouples that have useful measurering temperature up to 1500°C. It is found that reproducibility in the temperature of Melting Plateau is 0.1°C and that that of Freezing Plateau is 0.1°C.