SICE Annual Conference Program and Abstracts
SICE Annual Conference 2002
Conference information

NIC EUTECTIC POINT
Motoya Imai
Author information
CONFERENCE PROCEEDINGS FREE ACCESS

Pages 227

Details
Abstract

NiC eutectic point (approximately 1329°C) is studied as a fixed point for the calibration of Pt/Pd thermocouples that have useful measurering temperature up to 1500°C. It is found that reproducibility in the temperature of Melting Plateau is 0.1°C and that that of Freezing Plateau is 0.1°C.

Content from these authors
© 2002 SICE
Previous article Next article
feedback
Top