SICE Annual Conference Program and Abstracts
SICE Annual Conference 2002
Conference information

Infrared Radiation Thermometry Near Room Temperature at NMIJ
Juntaro IshiiYuka KanekoYukiko ShimizuFumihiro Sakuma
Author information
CONFERENCE PROCEEDINGS FREE ACCESS

Pages 457

Details
Abstract
Facilities and calibration techniques for infrared radiation thermometry in the low temperature range have been developed at the NMIJ. The estimated uncertainties (k = 2) of the radiance temperature of the reference blackbody sources are better than 60 mK in the temperature range from -20 °C to 160 °C. A compact infrared radiation thermometer has been developed as a transfer instrument. The uncertainty of calibration of the industrial standard blackbodies has been estimated to be smaller than 0.1 °C.
Content from these authors
© 2002 SICE
Previous article Next article
feedback
Top