SICE Annual Conference Program and Abstracts
SICE Annual Conference 2002
Conference information

Development of a spectral response calibration system using a wavelength-tunable laser and an integrating sphere
Yasuji Yamamoto
Author information
CONFERENCE PROCEEDINGS FREE ACCESS

Pages 460

Details
Abstract
A spectral response calibration system was developed using CW wavelength-tunable lasers as the monochromatic light source and an integrating sphere to provide a large, spatially uniform and depolarized light source. The basic characteristics of this system were evaluated. This system enabled to calibrate the spectral response of an optical sensor with a large aperture.
Content from these authors
© 2002 SICE
Previous article Next article
feedback
Top