Bulletin of the Society of Sea Water Science, Japan
Online ISSN : 2185-9213
Print ISSN : 0369-4550
ISSN-L : 0369-4550
Colorimetric Determination of Magnesium Ion with Xylidylblue
Studies on the Automatic Analysis of Salt Manufacturing Samples (Part 4)
Takahiro KANOUShizuo SUGITA
Author information
JOURNALS FREE ACCESS

1981 Volume 35 Issue 1 Pages 28-32

Details
Abstract

In order to determine automatically the magnesium ion of salt manufacturing samples, the authors made an investigation on Technicon's auto-analyzer by the colorimetric method with xylidylblue-II reagent, and then examined the effects of ethanol concentration of sample solution and of coexisting ions. The following were the results obtained from this study: 1) The reaction between magnesium ion and xylidylblue-II reagent is influenced by the alcoholic concentration. Therefore, it is necessary to control the alcoholic concentration to 54%(the final content). 2) Interference by the coexisting ions such as Na+ and Ca2+can be eliminated by the addition of G-EDTA reagent.
In this method, a linear relationship is obtained between the absorbance and the concentration of magnesium ion in the range of 0-20mg/l.
This method has the same accuracy as the conventional EDTA method, and can analyze continuously 15-20 samples per hour.

Information related to the author
© The Society of Sea Water Science, Japan
Previous article Next article
feedback
Top