Tetsu-to-Hagane
Online ISSN : 1883-2954
Print ISSN : 0021-1575
ISSN-L : 0021-1575
Regular Article
Determination of Ni in Fe-based Alloys using Low Pressure Laser-Induced Argon Plasma Emission Spectrometry
Chikage AbeKazuaki Wagatsuma
Author information
JOURNAL OPEN ACCESS

2012 Volume 98 Issue 2 Pages 63-68

Details
Abstract
In order to evaluate the analytical precision in low pressure laser-induced argon plasma emission spectrometry (LIPS), we conducted the determination of Ni in Fe-based alloys to consider the most suitable measurement condition, in particular, a gas pressure of argon and several measuring parameters for time-resolved observation. The selection of emission lines was an important factor to obtain a liner calibration curve depending on the concentration of Ni. The precision was generally inferior to analytical results in XRF and ICP-OES. However, the analytical value of Ni was obtained with good reliability, by averaging during replicate measurements. It was further shown the Ni content in Fe-Ni-X alloys would be determined by using calibration curves estimated from binary Fe-Ni alloys.
Content from these authors
© 2012 The Iron and Steel Institute of Japan

This article is licensed under a Creative Commons [Attribution-NonCommercial-NoDerivatives 4.0 International] license.
https://creativecommons.org/licenses/by-nc-nd/4.0/