Transactions of the Materials Research Society of Japan
Online ISSN : 2188-1650
Print ISSN : 1382-3469
ISSN-L : 1382-3469
Regular Papers
Comparison between Zn-modification and groove-formation towards better osseointegration of implant
Yaozhao LiOsamu Yamamoto
Author information
JOURNAL FREE ACCESS

2014 Volume 39 Issue 3 Pages 345-348

Details
Abstract
Ti implants with two-type surfaces, regular micro-groove and Zn-modification, were prepared in order to assess the effect of the implant surface on the osseointegration between bone and implant. Biomechanical push-out testing was performed to evaluate bone-implant interfacial shear strength at 4, 8 and 12 weeks after implantation to rabbit femora. The results revealed that there was no significant effect of the regular micro-groove on the bone fixation of the implants. On the other hand, Zn-modified implant was found to be effective for producing a stronger bone fixation than those with the regular micro-groove.
Content from these authors
© 2014 The Materials Research Society of Japan
Previous article Next article
feedback
Top