Risk of Blood Glucose Measurements with Handheld Meters in Anemia Patient
Fumiatsu YakushijiHiroshi FujitaMutsuko YasudaKen TaniguchiKazuhiko FujikiAkira OowadaNobu AkiyamaHiroshi SuzukiRitsuko JokyuYoshiyasu TerayamaMasako Shimojo
Author information
JOURNAL FREE ACCESS

2008 Volume 51 Issue 9 Pages 837-843

Details
Abstract
Backgrounds: Hand-held blood sugar meter operation is reported to have become abnormal at experimentally artificial hematocrit. We experimentally and clinically evaluated measurement with especially low hematocrit.
Methods: We first prepared three affiliated samples-artificial Ht adjusted blood (1), whole blood from anemic patients (2), and serial samples of blood from an anemic patient with diabetes (3)-and calculated hand-held blood sugar meters for their bias.
Results: We found clinically that sugar concentration discrepancies using hand-held meters are significant in anemic patients.
Conclusion: Although convenient to use, hand-held blood sugar meters present risks in measurement, especially in anemic patients. The approved Ht blood sampling ranges of these meters must thus be taken into account in their use.