Report of the Committee
Recommendation of the Use of EDTA-containing Tubes for the Measurement of HbA1c Using a Layer of Erythrocytes Obtained after Centrifugation
[in Japanese]Katsuhiko KuwaMikiko OkahashiAsako SatoMitsuyoshi NambaShiro MaedaMasami MurakamiShu MeguroYoshihiko Nishio
Author information
JOURNAL FREE ACCESS

2021 Volume 64 Issue 5 Pages 336-339

Details
Abstract

Transportation of whole blood samples maintained in cold conditions for up to 48 hours out of the laboratory may lead to hemolysis. This consequently results in a lower measured HbA1c value when a layer of erythrocytes obtained by centrifugation of the hemolytic sample is used for an analysis, including analyses by some HPLC-, enzymatic- and immuno-assay systems. The use of NaF-containing tubes that are suitable for the measurement of glucose concentration, makes hemolysis easy. Thus, the committee on Standardization of Laboratory Testing Related to Diabetes Mellitus of the Japan Diabetes Society recommends the use of EDTA- containing tubes for the measurement of HbA1c using a layer of erythrocytes harvested by centrifugation.