Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Flow-Injection Assay of Superoxide Dismutase Based on the Reduction of Highly Water-Soluble Tetrazolium
Hiroyuki UKEDAAshok K. SARKERDaisuke KAWANAMasayoshi SAWAMURA
Author information
JOURNAL FREE ACCESS

1999 Volume 15 Issue 4 Pages 353-357

Details
Abstract

A flow-injection analysis (FIA) for superoxide dismutase (SOD) activity was developed based on the use of tetrazolium salt, WST-1 (4-[3-(4-iodophenyl)-2-(4-nitrophenyl)-2H-5-tetrazolio]-1,3-benzene disulfonate sodium salt) and an enzyme reactor packed with Sepharose 4B on which xanthine oxidase (XO) and catalase were co-immobilized. WST-1 is highly water-soluble, and no adhesion of the reduced form to the FIA line was observed during continuous operation for 3 months. As optimized conditions, a sample (9 vol) was mixed with a reagent solution (1 vol) containing 3 mM hypoxanthine and 2 mM WST-1, and the mixed solution (20 μl) was injected into a carrier stream of 50 mM carbonate buffer (pH 10.2) at a flow rate of 0.4 ml/min. Under the conditions, the concentration of the SOD preparation giving 50% inhibition (IC50) was 2.7 μg/ml and the sampling frequency was 30 samples/h. When the SOD activity in erythrocytes of rats was determined by the present FIA method, the values were linearly related to those obtained by the conventional nitroblue tetrazolium (NBT) assay (r=0.975; n=18). The enzyme reactor was stable for at least 200 repetitive injections.

Content from these authors
© 1999 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top