Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Accuracy of Total Reflection X-Ray Fluorescence Spectrometry near the Detection Limit
Yoshihiro MORIKazuhiko KUBOTAKengo SHIMANOETadashi SAKON
Author information
JOURNAL FREE ACCESS

1998 Volume 14 Issue 2 Pages 275-280

Details
Abstract

Total reflection X-ray fluorescence spectrometry (TXRF) is one of the most common tools to analyze metal contaminants on silicon wafers and other substrate surfaces. Although the detection limit of commercial TXRF was improved to the concentration region of 109 atoms cm-2, its accuracy at low concentration was not yet clarified until now. In this paper, we examine the accuracy of the quantitative analysis by TXRF under 1011 atoms cm-2 for Fe, Ni, Cu and Zn. In particular, four factors (standard sample, reference analyzing method, compensation of spurious peak and peak separation) are considered. Under a controlled condition, the accuracy is within 20% as compared with atomic absorption spectropho-tometry (AAS).

Content from these authors
© 1998 by The Japan Society for Analytical Chemistry