Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Large Aliquot Injection Approach in the Electrothermal AAS Determination of Arsenic in Environmental Water Samples
Shoji IMAIKenichiro FUJIKAWAAkira YONETANINobuaki OGAWAYoichi KIKUCHI
Author information
JOURNAL FREE ACCESS

2000 Volume 16 Issue 2 Pages 163-167

Details
Abstract
Determination of As by direct electrothermal atomization atomic absorption spectrometry with a pyropytically-coated graphite furnace treated with W by a single drop treatment technique was reported. The treated PG furnace made it possible to introduce a 100 µl aliquot of sample solution into the furnace with a high precision (RSD <4%). Use of a Pd-modifier made it possible to enhance the precision (RSD <3%) and the sensitivity. A good linearity (correlation factor >0.99) between integrated absorbance and injection volume was observed in the range 0 - 100 µl. The calculated detection limit and characteristic mass (sensitivity) for integrated absorbance were 22 ng l-1 and 16 pg and for maximum absorbance were 45 ng l-1 and 13 pg for a 100 µl injection with the Pd modifier at the optimum ashing temperature of 800°C. An interference study for environmental matrices was also carried out. The recoveries of 2 and/or 4 µg l-1 of As added to various water samples, such as snow, cloud, river and tap waters, were 97 - 104%.
Content from these authors
© 2000 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top