Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Matrix Effect on the Analysis of Oligonucleotides by Using a Mass Spectrometer with a Sonic Spray Ionization Source
Min HUANGAtsumu HIRABAYASHIAkihiko OKUMURAYukiko HIRABAYASHI
Author information
JOURNAL FREE ACCESS

2001 Volume 17 Issue 10 Pages 1179-1182

Details
Abstract

Matrix or impurities remaining in a DNA sample solution after various sample treatment procedures may influence a subsequent DNA analysis. In this work, several matrices were investigated concerning their effects on the analysis of oligonucleotide by using an ion-trap mass spectrometer equipped with a sonic spray ionization source. Inorganic salts of sodium chloride and magnesium chloride depressed the signal intensity by about 50% when the content of the salts was about 10 µM. dNTPs and Taq showed more severe depression on the oligonucleotide. However, Tris, or (hydroxymethyl)aminomethane, intensified the signal intensity, if its content was within an appropriate range. When the content of Tris was about 500 µM, the signal intensity was enhanced by factors of 3 and 5 for the 6-mer and the 20-mer oligonucleotides, respectively. With the existence of Tris, matrix effects from the inorganic salts, dNTPs and Taq were reduced.

Content from these authors
© 2001 by The Japan Society for Analytical Chemistry
Previous article Next article