Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
A Nitrate-Selective Electrode Based on Tris(2,2′,2″-salicylideneimino) triethylamine
Nazife ASLAN (YILMAZ)Adnan KENAROrhan ATAKOLEsma KILIC
Author information
JOURNAL FREE ACCESS

2001 Volume 17 Issue 11 Pages 1269-1272

Details
Abstract
A new nitrate-selective liquid-membrane electrode based on the tris(2,2′,2″-salicylideneimino)triethylamine-iron(III) complex has been developed. This electrode exhibits a linear Nernstian response over the range 10-1 - 10-4 M of nitrate, with a slope of 54.3 ± 0.9 mV per p[NO3]. The effects of the pH and the liquid-membrane composition were also investigated. The lifetime of the electrode is at least one month. The selectivity coefficients for ten monovalent ions were calculated. The preparation procedure of the electrode is very easy and inexpensive. Moreover, the proposed electrode was applied for the determination of nitrate in fertilizers.
Content from these authors
© 2001 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top