Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Simple and Sensitive Determination of Citrinin in Monascus by GC-Selected Ion Monitoring Mass Spectrometry
Pin-Yen SHUCheng-Huang LIN
Author information
JOURNAL FREE ACCESS

2002 Volume 18 Issue 3 Pages 283-287

Details
Abstract

A new method for the qualitative and quantitative analysis of citrinin in Monascus by gas-chromatography-selected ion monitoring (SIM) mass spectrometry has been developed. GC separation of citrinin in Monascus extract was achieved without the need for chemical derivatization, and could be detected as a single peak when the SIM mode selected 5 prominent fragmentations (m/z of 220, 205, 177, 105 and 91). The quantitative detection limit for citrinin was ∼1 ppb. Finally, the GC-separated analyte from Monascus extract, at a retention time of 10.89 min, was examined by the method of pattern recognition by comparison with a citrinin standard. The results show that the 2 compounds had a 94% similarity when the SIM mode was used.

Content from these authors
© 2002 by The Japan Society for Analytical Chemistry
Previous article Next article