2003 Volume 19 Issue 1 Pages 129-132
We constructed an assembled-type MgF2 cell that can function under a high vacuum (10-4 Pa), and is capable of measuring the vacuum-ultraviolet circular dichroism (VUVCD) spectrum in a wavelength region down to 140 nm for aqueous solutions. Its path length can be adjusted by various spacers in the range from 1.3 to 50 μm. The temperature-control unit of the cell was also constructed with a Peltier thermoelectric element to keep the temperature of a sample within an accuracy of ±1°C in the temperature range from −30 to 70°C. The optical cell and the temperature-control unit were confirmed to have good performance by monitoring the VUVCD spectra of ammonium d-camphor-10-sulfonate and myoglobin aqueous solutions. This cell is available not only for VUVCD spectroscopy, but also for vacuum-ultraviolet absorption measurements.