2003 Volume 19 Issue 12 Pages 1671-1674
A novel modified glassy carbon electrode containing Ni(Me2(CH3CO)2[14]tetraenoN4) complex was used as an electrocatalytic sensor for the determination of L-ascorbic acid in pH = 6.6. The peak potential shifted to negative by 205 mV compared with that for a bare electrode in cyclic voltammograms. The calibration curve was linear up to 6.2 × 10-3 M with a detection limit 3.1 × 10-7 M and an RSD% better than 2.47%. This newly modified electrode was applied to commercial pharmaceutical tablets, injections and foods. The obtained results were identical to those obtained by the classical 2,6-dichlorophenolindophenol method.