Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Notes
Determination of Inorganic Anions in Natural Water by Microcolumn Ion Chromatography with On-column Enrichment
Lee Wah LIMJi-Ye JINToyohide TAKEUCHI
Author information
JOURNAL FREE ACCESS

2003 Volume 19 Issue 3 Pages 447-449

Details
Abstract
An on-column enrichment method was developed for the rapid determination of inorganic anions in natural water. The system was assembled from a syringe pump, a six-port switching valve with a sample-enrichment loop, a separation column and a UV detector. The enrichment efficiency of the system was tested by using inorganic anions as samples. The limits of detection were between 0.6 and 7.7 μg/L. The system was applied to the determination of anions in river and pond-water samples.
Content from these authors
© 2003 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback