Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Sensitive and Simple Cloud-point Preconcentration Atomic Absorption Spectrometry: Application to the Determination of Cobalt in Urine Samples
Jamshid L. MANZOORIGhasem KARIM-NEZHAD
Author information
JOURNAL FREE ACCESS

2003 Volume 19 Issue 4 Pages 579-583

Details
Abstract

A new approach for developing a cloud-point extraction-flame atomic absorption spectrometric method has been described and used for the determination of cobalt. In this approach, water was removed from the final diluted surfactant rich phase obtained in cloud-point extraction procedure. The results indicated that removing water from this phase increased the enhancement factor by 4-fold. 1-(2-Pyridylazo)-2-naphthol (PAN) and octylphenoxypolyethoxyethanol (Triton X-114) were used as a hydrophobic ligand and a nonionic surfactant, respectively. The chemical variables affecting the preconcentration step were optimized. The effect of the water concentration in the final diluted methanolic surfactant solution on the analytical signal was investigated. The results showed that the analytical signal decreased by 30% and 52% in 15% and 25% water concentrations in methanol, respectively. An enhancement factor of 115 was obtained for cobalt extracted from only 10 ml of a sample. The detection limit obtained under the optimal condition was 0.38 μg l-1. The proposed procedure was applied to the determination of cobalt in urine samples.

Content from these authors
© 2003 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top