Flow-through Micro Sensor Using Immobilized Peroxidase with Chemiluminometric FIA System for Detemining Hydrogen Peroxide
Nobutoshi KIBAToyonari TOKIZAWASayaka KATOMasaki TACHIBANAKazue TANIHitoshi KOIZUMIMasaharu EDOEiichi YONEZAWA
Author information
JOURNAL FREE ACCESS

2003 Volume 19 Issue 6 Pages 823-827

Details
Abstract

A micromachined flow cell (overall size; 25 × 25 × 1 mm3) was designed for the fast determination of hydrogen peroxide, based on a luminol-H2O2 chemiluminescence reaction catalyzed by immobilized peroxidase (POD). The flow cell consisted of a sandwich of anisotropically etched silicon and glass chips and contained a spiral channel (20 turns, 50 cm long, 150 μm wide, 20 μm depth, channel volume 1.4 μl) and two holes (1 mm diameter). POD was covalently immobilized with 3-(trimethoxysilyl)propyldietylenetriamine and glutaraldehyde on the inner surface of the channel. The chip was placed in front of a window of a photomultiplier tube and used as a flow cell in a single-line flow-injection analysis system using a luminol solution as a carrier solution. The sample volume for one measurement was 0.2 μl. The maximal sampling rate was 315 h-1 at a carrier solution flow rate of 10 μl min-1. A calibration graph for H2O2 was linear for 5 nM - 5 μM; the detection limit (signal-to-noise = 3) was 1 nM (7 fg in 0.2 μl injection). The H2O2 concentration in rainwater was determined using this sensor system.