Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Detection of Chlorobenzene Derivatives Using Vacuum Ultraviolet Ionization Time-of-Flight Mass Spectrometry
Kenichi TONOKURATomohisa NAKAMURAMitsuo KOSHI
Author information
JOURNAL FREE ACCESS

2003 Volume 19 Issue 8 Pages 1109-1113

Details
Abstract

Vacuum ultraviolet single-photon ionization time-of-flight mass spectrometry (VUV-SPI-TOFMS) has been applied for the detection of chlorobenzene, o-dichlorobenzene, and o-chlorophenol as surrogates for polychlorinated dibenzo-p-dioxine/furans (PCDD/F). The photoionization mass spectra of these compounds appear to be fragmentation free in the ionization processes by the VUV-SPI at 10.2 eV (121.6 nm). Quantum chemical calculations support no fragmentation in the photoionization of chlorobenzene derivatives at around 10 eV. The absolute photoionization cross-sections of chlorobenzene, o-dichlorobenzene, and o-chlorophenol were estimated at 10.2 eV. The photoionization cross-section is an important parameter in the detection of chlorobenzene derivatives by the single-photon ionization technique. The detection limit for chlorobenzene is on the order of tenth parts-per-billion volume (ppbv) in the present experimental setup.

Content from these authors
© 2003 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top