Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Notes
Polydimethylsiloxane Connection for Quartz Microchips in a High-Pressure System
Yukihiro SHINTANIKeiji HIRAKOMasanori MOTOKAWAYoshihiko TAKANOMasahiro FURUNOHiroyoshi MINAKUCHIMitsuyoshi UEDA
Author information
JOURNAL FREE ACCESS

2004 Volume 20 Issue 12 Pages 1721-1723

Details
Abstract

An interconnecting technique, the “DMS connection method”, for quartz microchips in a high-pressure system is presented. The connection between quartz microchips is an essential technology for modular microfluidic devices, such as microchip-HPLC. PDMS was applied to the seal material, being spread on the seal side of the chips, and set into the metal housing. The characteristics of the PDMS connection method concerning pressure resistance and the extension of the peak were examined. The experimental results showed a good seal at 5 MPa, which seem to be sufficient for realizing microchip-HPLC utilizing a monolithic silica capillary column as a separation medium. The influence of the extra column effect on chromatographic separation was almost the same as in the case using a commercial union fitting. In addition, the PDMS connection enabled the detachability of chip-based modules with user-friendliness. Our experimental findings suggest that the novel PDMS connection method can possibly be applied as a generic technology in high-pressure µTAS.

Content from these authors
© 2004 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback