Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Coated Wire Silver-Ion Selective Electrode Based on a N,N′-Bis(2-thienylmethylene)-1,2-diaminobenzene
Hamid R. ZAREMasoud SALAVATI-NIASSARYFarkhondeh MEMARZADEHMohammad MAZLOUMNavid NASIRIZADEH
Author information
JOURNAL FREE ACCESS

2004 Volume 20 Issue 5 Pages 815-819

Details
Abstract

A novel membrane coated platinum-wire electrode (MCPWE) based on N,N′-bis(2-thienylmethylene)-1,2-diaminobenzene (BTMD) for highly selective determination of Ag+ ion has been developed. The influences of membrane composition and pH on the potentiometric responses of electrode were investigated. The potentiometric responses are independent of the pH of the test solution in the range of 5.0 - 9.0. The electrode shows a linear response for Ag+ ion over the concentration range of 1.0 × 10-6 to 1.0 × 10-1 M with a lower detection limit of 6.0 × 10-7 M. The electrode possesses a Nernstian slope of 59.7 mV decade-1 and a fast response time of ≤ 17 s and can be used for at least 2 months without any observable deviation. The proposed electrode displayed very good selectivity for Ag+ ion with respect to NH4+ and alkali, alkaline earth and some common transition metal ions. The practical utility of the electrode has been demonstrated by its use as the indicator electrode in the potentiometric titration of an AgNO3 solution with a NaI solution and in determination of the silver content of a developed radiological film.

Content from these authors
© 2004 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback