Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Permanganate-based Chemiluminescence Analysis of Ferulic Acid Using Flow Injection
Li Na LINian Bing LIHong Qun LUO
Author information
JOURNAL FREE ACCESS

2005 Volume 21 Issue 8 Pages 963-966

Details
Abstract
A simple, sensitive and rapid flow-injection chemiluminescence method has been developed for the determination of ferulic acid based on the chemiluminescence reaction of ferulic acid with potassium permanganate in a nitric acid medium. A strong chemiluminescence signal was observed when ferulic acid was injected into an acidic potassium permanganate solution in a flow-cell. The present method allowed the determination of ferulic acid in the concentration range of 6.0 × 10-6 to 2.0 × 10-4 mol l-1; the detection limit (3 σ) for ferulic acid was 9.6 × 10-8 mol l-1. The relative standard deviation was 1.0% for 11 replicate analyses of 2.0 × 10-4 mol l-1 ferulic acid. The proposed method was applied to the determination of ferulic acid in real samples with satisfactory results. Moreover, the reaction mechanism of the chemiluminescence system was primarily considered.
Content from these authors
© 2005 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback