Pneumatic Flow Injection Analysis-Tandem Spectrometer System for Iron Speciation
Meissam NOROOZIFARMozhgan KHORASANI-MOTLAGHReza AKBARI
Author information
JOURNAL FREE ACCESS

2006 Volume 22 Issue 1 Pages 141-144

Details
Abstract
A pneumatic flow injection-tandem spectrometer system, without a delivery pump was used for the speciation of iron. In this system, the suction force of a pneumatic nebulizer of a flame absorption spectrometer was used for solution delivery through the manifold. The Fe(III) and total Fe concentrations were determined using thiocyanate ion in a UV-Vis spectrometer and a FAAS, respectively. The Fe(II) was determined by the difference. The calibration curves were linear up to 18 µg mL-1 and 25 µg mL-1 with detection limits of 0.09 µg mL-1 and 0.07 µg mL-1 for Fe(III) and Fe(II), respectively. The mid-range precision and accuracy were <2.5% and ±3% for the two species, respectively, at a sampling rate of 120 h-1. This system was applied for the determination of Fe(III) and Fe(II) in industrial water, natural water and spiked samples.