Simple Spectrophotometric Flow Injection System with an In-valve Minicolumn for Enhancement during the Determination of Chromium(III) Using EDTA
Somchai LAPANANTNOPPAKHUNSupharphorn KASUWASLucksagoon GANRANOOJaroon JAKMUNEEKate GRUDPAN
Author information
JOURNAL FREE ACCESS

2006 Volume 22 Issue 1 Pages 153-155

Details
Abstract

A simple spectrophotometric flow injection (FI) procedure for the determination of Cr(III) using ethylenediaminetetraacetic acid (EDTA) was developed. An FI system with a column packed with Amberlite IR-120(H) was employed for sample pretreatment. This leads to the possibility of a single standard calibration. A linear calibration in a range of 10 - 27 µg Cr(III) was obtained with a detection limit of 1 µg Cr(III) and RSD of 2% (18 µg Cr(III), n = 12). The proposed procedure was applied for determination of Cr(III) in leachate and dietary supplement samples. The results agreed with those obtained by the standard methods.

Content from these authors
© 2006 by The Japan Society for Analytical Chemistry