2006 Volume 22 Issue 5 Pages 673-678
A new sulfate-selective electrode based on the complex N,N′-bis(2-amino-1-oxo-phenelenyl)phenylenediamine copper(II) (CuL) as the membrane carrier was developed. The electrode exhibited a good Nernstian slope of -29.5 ± 0.5 mV/decade and a linear range of 1.0 × 10-7 - 1.0 × 10-1 M for sulfate. The limit of detection was 1.0 × 10-8 M. It has a fast response time of 10 s and can be used for more than three months. The selective coefficients were determined by the fixed interference method (FIM). The electrode could be used in the pH range 3.5 - 8.0. It was employed as an indicator electrode for direct determination of sulfate in pharmacy and cement samples.