Abstract
The possibility was investigated by using 1-phenyl-3-methyl-4-benzoyl-5-pyrazolone (PMBP) as the chelating reagent for separation and preconcentration of manganese(II) by cloud point extraction (CPE) and subsequent determination by flame atomic absorption spectrometry (FAAS). The effects of experimental conditions such as pH, concentration of chelating agent and surfactant, equilibration temperature and time on cloud point extraction were studied. Under the optimum conditions, preconcentration of 10 ml of sample solution permitted the detection of 1.45 ng mL-1 of manganese with an enrichment factor of 20. The proposed method was applied to the determination of trace manganese in water samples with satisfactory results.