Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Rapid Communications
Stopped-in-Loop Flow Analysis of Trace Vanadium in Water
Norio TESHIMAShinsuke OHNOTadao SAKAI
Author information
JOURNAL FREE ACCESS

2007 Volume 23 Issue 1 Pages 1-2

Details
Abstract
The new concept of stopped-in-loop flow analysis (SIL-FA) is proposed, and an SIL-FA method for the catalytic determination of vanadium is demonstrated. In an SIL format, a sample solution merges with reagent(s), and the well-mixed solution is loaded into a loop. The solution in the loop is separated by a six-way switching valve from the main stream. While the reaction proceeds in the stationary loop, the SIL-FA system does not need to establish a baseline continuously. This leads to a reduction in reagent consumption and waste generation compared with traditional flow injection analysis.
Content from these authors
© 2007 by The Japan Society for Analytical Chemistry
Next article