Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
On-line Preconcentration Method for the Determination of Trace Metals in Water Samples Using a Fully Automated Pretreatment System Coupled with ICP-AES
Rosi K. KATARINANarong LENGHORShoji MOTOMIZU
Author information
JOURNAL FREE ACCESS

2007 Volume 23 Issue 3 Pages 343-350

Details
Abstract

An automated on-line sample-preparation method using a computer-controlled pretreatment system (Auto-Pret AES system) coupled with ICP-AES was developed. In this work, an iminodiacetate chelating resin, packed in a mini-column and installed in the system was employed for the collection/concentration of 13 trace metals, including such toxic metals as Be, Cd, Cr, Cu and Pb. The limits of detection of the proposed method for trace metals were in the range of 0.001 (Be) - 0.18 (Pb) ng mL-1. The enrichment factors for metal ions were about 19 times, when 5 mL of samples were used. The sample throughput was 11 h-1. The accuracy and the precision of the method were evaluated using river-water reference materials, SLRS-4 from NRCC, JSAC 0301-1 and JSAC 0302 from the Japan Society for Analytical Chemistry. The proposed method can be favorably applied to the collection/concentration of trace metals in natural water samples.

Content from these authors
© 2007 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback