Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Rapid and Sensitive HPLC Method for the Simultaneous Determination of Paraquat and Diquat in Human Serum
Shuuji HARANoriaki SASAKIDaizi TAKASEShouichi SHIOTSUKAKentaro OGATAKoujiro FUTAGAMIKazuo TAMURA
Author information
JOURNAL FREE ACCESS

2007 Volume 23 Issue 5 Pages 523-526

Details
Abstract
A rapid and sensitive HPLC method for the simultaneous determination of paraquat and diquat in human serum has been developed. After deproteinization of the serum with 10% trichloroacetic acid, the samples were separated on a reversed-phase column, and subsequently reduced to their radicals with alkaline sodium hydrosulfite solution. These radicals were monitored with a UV detector at 391 nm. This method permitted the reliable quantification of paraquat over linear ranges of 50 ng - 10 µg/ml and 100 ng - 10 µg/ml for diquat in human serum. The within- and between-day variations are lower than 2.3 and 2.2%, respectively. This technique was also utilized to determine the paraquat and diquat serum levels in a patient who had ingested herbicide (Prigrox L®) containing paraquat and diquat.
Content from these authors
© 2007 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top