Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
A Sensitive Chemiluminescence Method for the Determination of H2O2 in Exhaled Breath Condensate
Yufei HUZhujun ZHANGChunyan YANG
Author information
JOURNAL FREE ACCESS

2008 Volume 24 Issue 2 Pages 201-205

Details
Abstract

In this paper, a novel flow injection chemiluminescence (FI-CL) method is proposed for the determination of picomolar L-1 levels of hydrogen peroxide (H2O2) in exhaled breath condensate (EBC). This method is based on the oxidation of a low concentration of luminol (10-7 M) by H2O2 at a low concentration level (< 10-8 M) in an alkaline medium catalyzed by a complex, K5[Cu(HIO6)2] (DPC), which is not interfered by other metal ions or horseradish peroxidase (HRP). Under the optimum conditions, H2O2 was determined over the range of 1.0 × 10-10 to 1.0 × 10-8 mol L-1 with a detection limit of (3σ) of 4.1 × 10-11 mol L-1. The relative standard deviation (RSD) was 3.2% for 5 nmol L-1 H2O2 (n = 7). The proposed method offers the advantages of ultra-sensitivity, selectivity, simplicity and rapidity for H2O2 determination. It was successfully applied to directly determine trace amounts of H2O2 (nmol L-1) in human's EBC of both rheum and healthy volunteers. A statistically significant difference was found between patients with rheum (n = 11) and control subjects without rheum (n = 11).

Content from these authors
© 2008 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback
Top