Comparison of Enzymatic Recycling Electrodes for Measuring Aminophenol: Development of a Highly Sensitive Natriuretic Peptide Assay System
Yasuhiro MIEKeiko KOWATAYu HIRANOOsamu NIWAFumio MIZUTANI
Author information
JOURNAL FREE ACCESS

2008 Volume 24 Issue 5 Pages 577-582

Details
Abstract
Several redox enzymes were examined for enzymatic/electrochemical-recycling systems in order to measure p-aminophenol (PAP) with high sensitivity. Glucose oxidase (GOD) and diaphorase (DI) worked well as catalysts for recycling electrode systems: these enzymes effectively reduced p-iminoquinone (PIQ), the electrochemically-oxidized form of PAP, and caused an enhancement in the electrochemical signals (anodic currents in the voltammogram and amperogram) by ∼ 100 fold. The lower detection limits for PAP were estimated to be 50 nM with the GOD system and 2 nM with the DI system. We combined the enzymatic-recycling electrode using DI with an enzyme immunoassay system to measure atrial natriuretic peptide (ANP), an important marker peptide hormone involved in heart diseases. ANPs from serum samples at ppt-levels were determined appropriately using the present assay system.