Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Determination of Benzophenones in River-water Samples Using Drop-based Liquid Phase Microextraction Coupled with Gas Chromatography/Mass Spectrometry
Noriya OKANOUCHIHidehiro HONDARie ITOMigaku KAWAGUCHIKoichi SAITOHiroyuki NAKAZAWA
Author information
JOURNAL FREE ACCESS

2008 Volume 24 Issue 5 Pages 627-630

Details
Abstract

In this study, a fast, simple and highly sensitive method that employs liquid phase microextraction (LPME)-GC/MS was developed to analyze trace benzophenones (BPs) in river-water samples. The tip of a 10-µl microsyringe filled with toluene (3 µl) was inserted into 2 ml of a river-water sample, and fixed at 5 mm below the water surface of the sample. A toluene droplet was made on the tip of the syringe, and extraction was conducted while agitating at 500 rpm for 15 min. After extraction, 2.0 µl of the extract was put into the syringe again, and directly introduced to GC/MS. The limits of detection (S/N = 3) and quantification (S/N >10) of BPs were 10 and 50 pg ml-1, respectively. The results of a recovery test ranged over 93.3 - 101.1% (RSD, less than 10%; n = 6). The results of BPs determinations in the river-water samples showed that BPs (ND - 68.9 pg ml-1) were detected.

Content from these authors
© 2008 by The Japan Society for Analytical Chemistry
Previous article Next article