Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Organic-inorganic Composite Cation-exchanger: Poly-o-toluidine Zr(IV) Phosphate-based Ion-selective Membrane Electrode for the Potentiometric Determination of Mercury
Asif Ali KHANINAMUDDINTabassum AKHTAR
Author information
JOURNAL FREE ACCESS

2008 Volume 24 Issue 7 Pages 881-887

Details
Abstract
A new heterogeneous precipitate of an organic-inorganic composite cation-exchanger poly-o-toluidine Zr(IV) phosphate was utilized for the preparation of a Hg(II) ion-sensitive membrane electrode for the determination of Hg(II) ions in real aqueous as well as in real samples. The electrode showed good potentiometric response characteristics, and displayed a linear log[Hg2+] versus EMF response over a wide concentration range of 1 × 10-1 - 1 × 10-6 M with a Nernstian slope of 30 mV per decade change in concentration with a detection limit of 1 × 10-6. The membrane electrode showed a very fast response time of 5 s and could be operated well in the pH range 2 - 8. The selectivity coefficients were determined by the mixed-solution method, and revealed that the electrode was selective in the presence of interfering cations; however most of these did not show significant interference in the concentration range of 1 × 10-1 - 1 × 10-4 M. The lifetime of the membrane electrode was observed to be 120 days. The analytical utility of this electrode was established by employing it as an indicator electrode in the potentiometric titrations of Hg2+ ions from a synthetic mixture as well as drain water.
Content from these authors
© 2008 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback