Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Notes
Analytical Method for Perchlorate in Water by Liquid Chromatography-Mass Spectrometry Using an Ion Exchange Column
Masahiro KAMOSHITAKoji KOSAKAMari ASAMIYukiko MATSUOKA
Author information
JOURNAL FREE ACCESS

2009 Volume 25 Issue 3 Pages 453-456

Details
Abstract
A practical analytical method was developed for the routine analysis of perchlorate in environmental and drinking-water samples by liquid chromatography-electrospray ionization mass spectrometry (LC/ESI-MS) using an anion exchange column. By using 18O-enriched perchlorate as an internal standard, the limits of quantification of perchlorate determined by tenfold of the signal-to-noise ratio and tenfold of the standard deviation were 0.1 and 0.03 µg L-1, respectively. The perchlorate concentrations in the raw and finished water samples from seven water purification plants were determined by LC/ESI-MS. Perchlorate was detected in 12 out of 13 samples, and the perchlorate concentrations in the samples were from 0.1 to 36.1 µg L-1.
Content from these authors
© 2009 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback