Analytical Sciences
Online ISSN : 1348-2246
Print ISSN : 0910-6340
ISSN-L : 0910-6340
Original Papers
Nano-Molar Level Hydrogen Peroxide Detection by Horseradish Peroxidase Adsorbed Cup-Stacked Carbon Nanotube Electrodes and Applications to L-Glutamate Detection
Tatsuo NODATadao UKAIToshio YAO
Author information
JOURNAL FREE ACCESS

2010 Volume 26 Issue 6 Pages 675-679

Details
Abstract
We have developed a simple fabrication method of a highly sensitive direct electron transfer-type electrochemical biosensor for hydrogen peroxide by use of cup-stacked carbon nanotubes (CSCNTs). The CSCNTs, formed by stacking of cup-shaped carbon units, has larger internal space and more hydrophilic edges, thanks to the presence of functional groups containing oxygen (e.g., –COOH, –OH), than multi-walled carbon nanotubes (MWCNTs). When the CSCNTs suspension was cast, the CSCNTs were dispersed homogeneously onto a glassy carbon (GC) electrode, and horseradish peroxidase (HRP) was immobilized firmly by physical adsorption without any chemical reactions. The flow injection analysis (FIA) system with the HRP/CSCNTs/GC electrode has superior sensitivity and stability to the HRP/MWCNTs/GC electrode. The detection limit was 0.75 nM (S/N = 3) and the activity was maintained over 85% for 21 days. Further, when the glutamate oxidase (GlOD)-immobilized reactor was set into the proposed FIA system, L-glutamate could be detected repeatedly with a detection limit of 1.2 nM (S/N = 3).
Content from these authors
© 2010 by The Japan Society for Analytical Chemistry
Previous article Next article
feedback